Product line
Bilirubinometers
NEO-BIL PLUS is an advanced and innovative version of traditional bilirubinometers. It is to analyze the ‘total bilirubin’ through a microcapillary tube in order to follow the new born icterus course. More than 20 years of experience in this specific field and that of physics-chemestry have led us to this achievement. Thanks to the use of the most advanced technology, frequent adjustments and continuous standard resetting are over. A precise sample measurement of both bilirubin and interferentials on the same point of the capillary is carried-out by a special reading system. The reading is carried out even on a small quantity of serum. No reagents are required. NEO BIL Plus is equipped with a built in printer for the results sequential printing.
Intended use | measurement of total bilirubin |
Sample | centrifuged whole blood |
Serum | minimum volume: 10 µl |
Reading cuvette | haeparinized haematocrit capillary |
Measure unit | mg/dl or µmol/l |
Measurement range | 0/30 mg/dl or 0/510 µmol/l |
Measure system | photometric |
Sample reading | on a single-point of the capillary |
Zero and checkings | completely automated |
Interferentials | automatically compensated |
Reading time | 3s approx. even on highly haemolized samples |
Reading inaccuracy | 3% in the range 5-30 mg/dl or 85-510 umol/l |
Detector | silicon photodiode |
Optical filters | 455 and 575 nm |
Programming | by membrane keypad |
Results | on LCD and printer |
Keypad | 17 buttons |
Display | 2 lines with 16 columns |
Printer | 20 columns with termal paper |
Data storage | last 1000 results |
Interface | USB port |
Other functions | built-in calendar and timer |
Dimensions | 25 x 22 x 9 cm |
Weight | Kg 1,6 |
Power requirements | 230/115Vac, 50-60Hz, 25W |
COMPLEMENTARY ITEMS | |
ACL-LP-0500 | lancettes |
ACL-PL-747500 | sealing plasticine |
ACL-CP-74931 | glass capillaries (75 mm) |
CRT-01-058 | thermal paper |