(+39) 0774 637070 Email: info@dasitaly.com Lun - Ven:   08.00 - 13.00   /   14:00 - 17:00

Technologies and Innovation

with DAS Instruments since 1977

ELISA - IFA AP22 IF ELITE

Product Line

Automatic Processor ELISA e IFA

AP22 IF ELITE is an automatic processor for ELISA and IFA methods. The instrument is capable of processing 2 ELISA microplates with 8 in-line tests or 16 IFA slides; in addition to the ELISA process, it is able to carry out sample predilution, sample and reagent dispensing, incubation and washing of IFA slides.

ELISA Technical Features
ELISA Dilution Cups 110 positions
Removable Trays modular trays for calibrators, controls, and reagents for 8 methods in line
Microplates 2 standard plates (12 x 8 wells) incubated at room temperature or from 36° to 45°C
Shaker Microplate shaker (optional)
Liquid Handling 1 needle for dilution and dispensing, the second for washing the wells of the slides
Sample Dispensing Time 11s with 2ml of wash
Microplate Washing 2 manifolds with 8 needles, one for dispensing and the other for aspiration; residual volume <3µl
Wash Buffer 2 wash buffers for ELISA microplates in external tanks with level sensors
Washing Accuracy CV <5% with 300 µl
Reading System 8 independent photometric channels for mono or bicromatic reading
Interference Filters 2 precision filters (450 and 630 nm); another 2 on request (4 filters max; 400-700 nm)
Reading Accuracy Reading from 0.000 to 3.000 OD; inaccuracy ± 1% from 0.000 to 1.500 OD ± 2% from 1.500 to 3.000 OD
IFA Technical Features
Liquid Handling 2 needles: one for dilution and dispensing, the other for slide washing
Sample Dispensing Time 11s with 2 ml of wash
Slide Process Time 30 minutes to dispense or wash 16 slides with 12 wells
Provided Trays Capacity for 10 tests in line; tray with 220 positions for dilutions; tray with 16 positions for IFA slides; tray for reagents for 18 controls, 9 conjugates, 1 wash solution for the needle, and two diluents for the samples
Liquid Sensor Able to capture 150µl
Wash Buffer 2 wash buffers for IFA slides in external tanks with level sensors
Washing Modes each well is washed individually to avoid contamination, continuous flow, or multi-well for specific washing effectiveness
Common Technical Features
Barcode Reader onboard barcode reader for samples
Sample Tray 91 sample tubes Ø13mm or 70 sample tubes Ø16mm that can be used in combination
Dilution Cups Blocks of 22 cups to facilitate loading
Syringe 1000µl volume with a resolution of 1µl
Dispensing from 8µl to 1000µl with a step of 1µl; inaccuracy ±1% at 100µl
Liquid Container External Wash and Waste tanks equipped with level sensors
Connections Possibility of LIS interfacing and loading worklist from Host computer. Possibility of connecting multiple instruments to a single PC.
Operating Systems Windows 7 or Windows 10
PC Minimum requirements: 250 MB free space on Hard Disk; CD-Rom reader; 2GB RAM; Intel Pentium Dual Core or equivalent, VGA 1024 x 768 16 million colors; 2 free USB 2.0 ports; printer.

Suggested requirements: 4GB RAM; Intel Core 2 Duo E6500 series or equivalent

Power Supply 230 / 115 VAC, 50-60 Hz, 350 W
Dimensions cm 62 x 83 x 72h
Weight 85 Kg