Neonatal bilirubin readers
NEO-BIL PLUS is the most advanced and innovative version of the Bilirubinometers that analyze total bilirubin on microcapillaries for monitoring neonatal jaundice. More than twenty years of experience in the specific field and in general in that of chemical-physical analysis has led us to achieve such an important result. The use of the most advanced technologies allows us to avoid frequent calibration and recalibration of the standard. The special reading system allows both bilirubin and interfering values to be read precisely on the same point of the capillary, even on samples with small quantities of serum. No reagents are required for the test. NEO BIL Plus is equipped with an internal printer for sequential printing of readings.
Intended Use | measurement of neonatal total bilirubin |
Sample | centrifuged whole blood |
Serum | minimum volume: 10 µl |
Reading Cuvettes | heparinized capillaries for hematocrit |
Units of Measurement | mg/dl or µmol/l |
Measurement Range | 0/30 mg/dl or 0/510 µmol/l |
Measurement System | photometric |
Sample Reading | at a single point of the capillary |
Zero and Controls | completely automatic |
Interferences | automatic subtraction |
Response Time | approximately 3s even on highly hemolyzed samples |
Reading Inaccuracy | 3% in the range of 5-30 mg/dl or 85-510 µmol/l |
Detector | silicon photodiode |
Optical Filters | 455 and 575 nm |
Programming | via keyboard |
Results | on display and on printer |
Keyboard | 5 keys |
Display | 2 lines with 16 columns |
Printer | 20 columns with thermal paper |
Data Storage | last 1000 readings |
Interface | USB port |
Other Functions | Clock/date and timer |
Dimensions | 25 x 22 x 9 cm |
Weight | Kg 1.6 |
Power Supply | VAC 110-260, 50-60Hz, 25W |
Additional Accessories | |
ACL-LP-0500 | fingerstick lancets |
ACL-PL-747500 | sealing clay |
ACL-CP-74931 | glass capillaries (75 mm) |
CRT-01-058 | thermal paper |