APE ELITE is the latest version of ELISA analyzers capable of managing 10 methods in line with 4 microplates at different temperatures. It is equipped with a barcode reader for sample slides on board; an internal microplate shaker is also available on request. Its “user friendly” approach allows the operator to start working with the instrument as soon as it is out of the packaging, no calibration or adjustment is required.
Barcode Reader | Integrated barcode reader for sample barcode reading |
Removable Trays | Modular trays for calibrators, controls, and reagents for 10 in-line methods |
Sample Tray | 169 sample tubes of Ø13 mm or 130 sample tubes of Ø16 mm; can be used in combination |
Dilution Tray | 264 dilution cups |
Sample Diluents | 4 on-board sample diluents |
Liquid Handling | 2 needles for dilution and dispensing of samples and controls |
Sample Dispensing Time | 6s (average) with 2 ml of wash |
Sample Dilution Time | 15s with 2 ml of wash |
Dispensing | CV < 0.1%, inaccuracy 1% at 100µl |
Aspiration | CV < 0.3%, inaccuracy 1% at 100µl; minimum volume 5µl, resolution 1µl |
Plate Washing | 2 manifolds with 8 needles, one for dispensing and the other for aspirating; residual volume < 3% |
Plate Washing Precision | CV < 5% at 300µl |
Microplates | 4 standard plates (12 x 8 wells) with independent incubation at room temperature or from 37° to 45° |
Syringes | 2 of 1000µl each, resolution 1µl |
Reading System | 8 independent photometric channels with the ability to read in mono and bichromatic |
Interference Filters | 2 precision filters (450nm and 630nm); two more on request (maximum 4 filters, 400nm-700nm) |
Reading Precision | reading from 0.000 to 3.000 OD, inaccuracy ± 2% |
Wash Buffer | 4 wash buffers for microplates in external tanks |
Liquid Containers | 8 external tanks with level sensors |
Operational Functions | Handling of 10 methods simultaneously. Definition of incubation times and volume to dispense for samples and reagents. Different incubation times for each method. Quick creation of new protocols. Worklist creation, patient report and result archiving. Saving calibration curve for each test performed. |
Connections | LIS interfacing, worklist import from Host Computer; a single PC can control multiple instruments in parallel |
PC Specifications | Minimum requirements: 250 MB free space on Hard Disk; CD-Rom reader; 2 GB RAM; Intel Pentium Dual Core or equivalent; VGA 1024 x 768 16 million colors, 1 free USB 2.0 port; printer |
Operating Systems | Windows 7 or Windows 10 |
Power Supply | 230 or 115 VAC, 50-60 Hz, 650 VA |
Dimensions | 113 x 77 x 75 h |
Weight | 120 Kg |